jueves, 28 de julio de 2011

EXPRESSION OF RESULTS AND REPORT OF TEST

Expression of results
Results are calculated using the following equation:







Where
G is the conductance, in siemens (S);
C is the capacity, in farads (F);
W is the angular frequency in radians / second (rad / s) = 2πf
F is the frequency used in Hertz (Hz).

NOTE - Assuming that the relative permittivity temperature measurement is known, the conductivity of the liquid can be calculated by the following equation:






Where








REPORT OF TEST
The report must include the following:
- Identification of the sample;
- test temperature;
- measured values of G and C;
- As calculated values

domingo, 10 de julio de 2011

SAMPLING,LABELLING AND PROCEDURES

SAMPLING
The insulating fluid samples should be taken by qualified personnel in accordance with IEC 60475. During his storage and transport, samples should be protected from direct light.

LABELLING
Insulating liquid samples must be properly labeled before being sent to the laboratory.
The following information is required:
- Customer or facility;
- Fluid identification (type and class);
- Identification of the equipment;
- The date and time of sampling;
- Temperature during sampling;
- sampling point;
- Other relevant information.

PROCEDURES
In order to obtain a meaningful measure of the dissipation factor is essential to follow exactly the rules with respect to:
- Careful cleaning of the test cell;
- Careful filling test cell and the careful handling of liquid samples and the very test cell.

Cleaning test cell
Procedure. Depending on the cleanliness of the test cell and the level of conductivity liquid to be analyzed, the cleaning of the test cell will be more or less sophisticated and more or take less time.
If the cleanliness of the cell is unknown, or if there is any doubt, you should have a cleaning process.
You can follow many cleaning procedures if they prove to be effective.

Checking the cleanliness of the empty cell. For a significant extent, it is necessary that
Cell electrical losses are much smaller empty the fluid to be measured.

Checking the cleanliness of the full cell for measurements at room temperature. If the cell trial is perfectly clean and fluid temperature is constant, and are independent of time. The measure can therefore be performed as soon as practically possible. In fact, this can be done in less 1 min. Moreover, a single measurement of a single sample is sufficient to obtain the correct value.
It may happen that, at constant temperature, conductivity(or so to increase or decrease over time, but no more than 2% at two minutes to fill the cell. In this case can be seen that the cell is sufficiently clean and the first measured value, ie a minute or so to fill the cell, can be registered.
Otherwise, it is recommended after cleaning the cell again, take a second sample of the same liquid and a second measure. It is considered the lowest value of the two, as recommended in IEC 60247.

Checking the full cell for measurements at temperatures above room temperature.
Before taking any action at high temperatures, make sure that the temperature of the liquid in the cell is constant. Except in cases in which the cell is perfectly clean, the measurement result depends on the way in which the cell and the liquid had been heated.
If the test cell is perfectly clean and the liquid temperature is constant, conductivity and are independent of time. The measure can thus be made immediately. In practice, the measure may do as soon as the temperature may be considered constant. Moreover, a single measurement of a single sample is enough to get the correct value.

It can happen, even if the liquid and the cell are at a constant temperature, conductivity (o de tan increase or decrease over time. This may be due to different causes: for example, heating to high temperatures can alter the composition of certain liquids, or modify the moisture content of the particles.
In practice, the temperature is not perfectly constant and their variations induce variations or conductivity,liquids varies more or less depending on the temperature nature of the liquid, typically up to 5% per degree centigrade. Therefore, the origin of variations.
If the test cell is not perfectly clean, the heating time affects the measured values, even the first, because the cell impurities dissolve in the liquid. The first measured value should, therefore, be discarded and the cell must be cleaned again.

Precautions when filling the test cell
When the cell is filled with liquid, you must ensure that the ambient atmosphere is, as far as possible, free of gas fumes that can be dissolved in the liquid.
The electrodes must be completely submerged in the liquid.


Test temperature
The measurement of conductivity and dissipation factor of a liquid can be made at any temperature.
Room temperature is recommended for its simplicity and time savings. Since the ambient temperature is essentially variable, you must remember a fixed value (eg 25 ° C).
There is nothing to stop the test is done at higher temperatures (eg 40 ° C 1 ° C, 90 ° C 1 º C or more).

Methods of heating
To make a measurement at high temperatures can be used several methods of heating. The time required to reach the test temperature depends on the method used and typically varies between 10 min and 60 min. If the test cell is not perfectly clean, the increase in conductivity due to the dissolution of impurities will depend on the duration of heating and the measured conductivity will depend, in turn, the heating method.
Thus, heating test cell as quickly as possible.
A suitable method to achieve this can be heated separately from the test cell and the liquid in a clean container. Another method is to rapidly heat the liquid in the test cell itself.

Measure
Fill the sample cell with avoiding contamination of the fluid or the cell (see 8.2).
Cleaning is checked as described in 8.1.3 or 8.1.4. If the cell is clean enough (see 8.1), is
Note the values of G and C.

sábado, 25 de junio de 2011

APPLIANCES

To perform this measurement method can be used specially constructed devices or constituted by individual items conveniently assembled. The block diagram shown in Figure 2 and elements described in the following sections illustrate appropriate equipment.

Block diagram of measuring apparatus









Keys

1. Test cell
2. Heating System
3. Square wave generator
4. Measurement chain
5. Meter
6. Recorder

Test Cell
The test cells of three terminals, designed according to the recommendations of IEC 60247, are generally appropriate for these measures.

You can use an additional type of cell in which there is no bridge, made by any solid insulating material, between the measuring electrodes, as shown in Figure 3. This cell type is often more accurate with highly insulating liquids.

Example of a cell test designed for highly insulating liquids













Keys

1. Cover
2. Internal electrode
3. External electrode
4. Stainless steel container
5. Thermocouple or thermometer to measure temperature
6. Electrical connections for BNC terminals

The typical distance between the inner and outer electrodes is 4 mm, the minimum distance should not be less than 1 mm.

The recommended material for the electrodes is stainless steel. As an example, the diameter of the outer electrode is 43 mm, the outer electrode is 51 mm, the length of the electrodes is 60 mm, and the diameter of the container Stainless steel is 65 mm.

This type of test cell is designed to minimize the effects of contamination of surfaces

Contact: although the contact surface is large, the relationship ÷ = "electrode surface" / "fluid volume" is relatively small (÷ = 2.6 cm-1) due to the large volume of liquid (v = 200 cm3).NOTE - It is recommended restricting the use of a given cell to a particular type of liquid.

Heating System
The heating system must be adequate to maintain the temperature of the measuring cell to within
±1 ºC of set point. This can be a forced draft oven or in an oil bath controlled thermostatically controlled and equipped with a bracket to hold the cell.
The heating system must be shielded electrical connections on the cell.

Square wave generator

The square wave generator must provide a quasi-rectangular voltage highly stable. The following features are appropriate:

- Amplitude: 10 V to 100 V;

- Frequency: 0.1 Hz to 1 Hz;

- Ripple: <1%;

- Rise time of tension: 1 ms to 100 ms.

Measurement Chain

The IR conduction current through the measuring cell, is measured in the second part of each half-wave and averages for a number of periods depends on the range of measures. The chain of measures gives the conductance G test cell.




As an example, the range of measured values ​​of conductance is 2 x 10-6 S a 2 x 10-14 S with a margin of error less than 2%.

The capacity C of the test cell is deduced from the measured current during the voltage rise. The values measurable capacity is between 10 pF and 1000 pF with an uncertainty of less than 1%.

As an example, for a liquid relative permittivity r = 2, a conductance value of 2 x 10-14 S gives

as ä = 0,8 x 10-6 a 50 Hz.

Twitter Delicious Facebook Digg Stumbleupon Favorites More

 
Design by Free WordPress Themes |distribuida por: Free blogger templates